LINDA

LINDA (X-ray LINe Detector with novel photon counting ASIC) is a pixel CdTe sensor with a built in fast algorithm (patent protected) to minimize the charge sharing among neighbouring pixels and thus maintains the spectral X-ray energy information per pixel. Each pixel has 6 thresholds to bin the spectral energy information into 6 levels. Each energy level has two counters one to read and one to acquire data so that the detector has zero dead time. One can see in fig.1 the kernel sensor of the of the X-ray line detector. The active size is 2.64 mm x 6.6 mm. When a number of such sensors are put side by side they will form a line detector that matches the scanner´s field of view.

Figure 1:
Left - Display the bare pixel ASIC for LINDA sensor with 8x20 pixel array and each pixel is 330um x 330um.
Right – Display the LINDA sensor, ASIC bonded to 1mm pixel CdTe detector and wire-bonded to dedicated readout electronics. The active area of the LINDA sensor is 2.64 mm x 6.6 mm

Figure 2:
Top - Display a pen.
Bottom – Display the radiography image of the pen using single LINDA sensor the pen was moved over the sensor
with the linear motor. As one can one can see the spring inside the pen and one can see that the plastic tube that
contains the ink is partially empty at the right hand side

 

Fig2 shows the results obtained using one LINDA sensor. One can see clearly the internal elements inside the pen. The image has not been subjected to image processing. Though the LINDA sensor is made for X-ray line scan machine, It can be tailored to fit the needs for the future Spectral Computed Tomography Scanner.
The LINDA sensor is unique because it has a fast algorithm to reduce the impact of the charge sharing process and thus preserve the spectral energy information detected. The preliminary report from the European Patent Office acknowledges the innovative steps in the charge summing mode for X-ray pixelated detector. At same time the LINDA sensor has relatively small pixel size (330um x 330um) for this type of application and with zero dead time. That is the sensor can acquire and send data simultaneously without loss of information. Given the LINDA sensor is an array of 8 (in direction of the moving belt) x 20 pixels this implies that object get imaged 8 times, by the 8 pixels as the object mover on the belt, and this provide a better image quality. This technique is called Time Delay Integration (TDI). With current version of LINDA sensor, one can achieve a scan speed of 20 m/min

 

The team of researcher will develop a 20cm X-ray line scanner based on LINDA sensor, which will be integrated into an X-ray machine provided by Multiscan Technologies SL.

This project has been co-financed by the European Union through the European Regional Development Fund (ERDF) and Agència de Gestió d’Ajuts Universitaris i de Recerca (AGAUR), and has been supported by the Secretaria d'Universitats i Recerca of Departament d'Empresa i Coneixement of Generalitat de Catalunya.

 

 

           